Screen 6: Fluid level sensor

The Grant digital ultrasonic bath is fitted with a fluid level sensor to ensure it is not
under-filled prior to or during the cycle. This ensures that the optimum cleaning
process can occur and potential damage to the tank is minimised.

The level sensor is a resistive device, so will not detect fluids with very low
conductivity such as pure deionised (DI) water.

The level sensor can be disabled for specialist cleaning applications where deionised
(DI) water is used.

Use the ON and OFF keys to set the Low Level function. When the desired option is
selected, use the MENU key to move to the next screen.

WARNING

Setting the fluid level sensor to OFF means that the

Grant digital

ultrasonic bath will

operate as if the bath were full of fluid.

It is therefore the operator‟s responsibility to ensure that the tank is filled with sufficient
water when the fluid level sensor is disabled.