Intek RheoVac 940 User Manual

Page 11

Advertising
background image

- 9 -

Figure 8. Output Connections and Set-up

TABLE I. Ten 4-20 mA Wire Terminal Assignment

OUTPUT VARIABLE

DESCRIPTION

CONFIGURATION PIN ASSIGNMENT

Active 4-20 mA

Passive 4-20 mA

!

+

!

+

ACTUAL VOLUME FLOW [ACFM]

A

B

B

A

TOTAL MASS FLOW [lbs/hr]

C

D

D

C

WATER VAPOR MASS FLOW [lbs/hr]

E

F

F

E

RheoVac PRESSURE ["Hg]

G

H

H

G

WATER VAPOR SPECIFIC VOLUME [cu. ft/lb]

I

J

J

I

WATER to AIR MASS RATIO

K

L

L

K

RELATIVE SATURATION [%]

M

N

N

M

PARTIAL PRESSURE, WATER ["Hg]

O

P

P

O

AIR IN-LEAK [SCFM]

Q

R

R

Q

RheoVac TEMPERATURE [°F]

S

T

T

S

NOT USED

U

U

U

U

Ð

RheoVac units have ten 4-20 mA outputs. These outputs can be configured collectively for either
passive or active transmitter. The units are shipped from the factory with the output jumpers in
the active position; i.e. the transmitter provides the current source. Figure 8 shows the locations
of the 4-20 mA select jumper, JP13 of the lower board — active position is shown.

.

CAUTION: changing the passive/active jumper changes the field wiring polarity and affects

all ten 4-20 mA channels. See Table I for wire terminal identifications.

Advertising