5.2.7 When a cycle completes, the display shows

Ed

the indication

light

RUNNING

turns off and the indication light

READY

turns on. At this time You may then open the chamber door and unload
the Instruments which are sterilized. (Fig 5-6) At the same time,
temperature indicate

prt

, it means that it is Performing print, whether

you Performing print, whether you Have install printer,the process can
not be omitted. But you needn't care about it. It will not effect the process
of opening the door and unloading the instruments. Print process will
Disappear automatically.

Attention: To ensure autoclave is safe, When water level low alarm sounds, Please be advised
add some water first. When you add the water, please don't forget to let condenser water out.

Caution: Always use the tray holder to load or unload the tray with instruments in
it, in order to avoid burning injuries.

If you didn't press

STEP

key during a cycle and the indication light is off, it means that

the machine is keep warm mode. The sterilization Chamber is kept being heated. This will
enable faster cycle that follows.

If you press

STEP

key before end of sterilization,

STEP

indication light goes on, it

means that the machine is last time cycle. The chamber starts to cool down completely before
the chamber door is open. It should normally be used during the very last cycle of the day.

Fig 5-6

Operation manual

Series Steam Sterilizers

10

5.2.8 If you decide to stop using the autoclave for a while, please turn the main power switch
off, or unplug the power cord should you not want to use the machine for a long time.

Caution: Never try to open the chamber door if and when the pressure display doesn't
show 0. It is suggested that chemical indicator or bio-indicator be used

for each and

every cycle to ensure the reliability of each sterilization cycle run.