Operation manual

Series Steam Sterilizers

16

After a long time usage, maybe some small impurity will deposit on the drain filter, it will
affect vacuum and drying. All the impurity is comes from the oil or dirt or dust of the
instrument which need sterilization, also maybe comes from the water. In order to prolong
the life of the drain filter, please note the following items:

2

rotate off the filter screw

3

pull out the filter from the

connection..

4

take off the blue rubber

which on the filter clean or
change a new one as you need.

5

after you clean or change

a new filter shim,plug in the
water filter back on the chamber
connection.Notice: please put
the blue rubber down and touch
the chamber and rotate tighten
the screw nut.

If draining filter plug up, you can solve according to fig 9-5.

-5

Fig9

1

find the water filter.

9.3 Change fuse (Fig 9-6)

1). Switch off the power.
2). Unscrew the fuse holder with a screw driver counter clockwise.
3). Pull the fuse holder out with your fingers.
4). Make sure to replace the cooked fuse with a correct fuse.
5). Put back the fuse holder and screw it on clockwise with ascrew driver.
6). You may use your hand to change the fuse.

-6

Fig9

Advertising