Measurement – LABEC V-11D User Manual

Page 10

Advertising
background image

- 7 -

Close the lid of the compartment and press

to calibrate Zero.

5) Calibrate 100%T

Pull the Reference in the light path, press

to calibrate 100%T。

6)

Print the result

Press the Button of

to print the test result.

Measurement

1. Measure the Absorbance

1) Press

to choose A mode (Fig.4-1)

Fig. 4-1

2)

Turn the wavelength knob till it displays the wavelength value you want.

3)

Pull the Reference in the light path, press

to set 100%T(Fig.4-2)

Fig. 4-2

4)

Pull the unknown concentration sample in the light path,then record the displayed value
(Fig. 4-3)。