LABEC V-11D User Manual

Page 5

Advertising
background image

- 2 -

Shows visible damage

Fails to perform the intended measurements

Has been subjected to prolonged storage under unfavorable conditions

Has been subjected to severe transport stresses

Working Principle:

The spectrophotometer consists of five parts:

1) Halogen or deuterium lamp to supply the light;

2) A Mono-chromator to isolate the wavelength of interest and eliminate the unwanted

second order radiation;

3) A sample compartment to accommodate the sample solution;

4) A detector to receive the transmitted light and convert it to an electrical signal;

5) A digital display to indicate absorbance or transmittance. The block diagram (Fig 1-1)

below illustrates the relationship between these parts.

Block diagram for the Spectrophotometer

Light Mono- Sample Detector

Display

Source

chromator

Compartment

Fig1-1

In your spectrophotometer, light from the lamp is focused on the entrance slit of the

monochromator where the collimating mirror directs the beam onto the grating. The

grating disperses the light beam to produce the spectrum, a portion of which is focused on

the exit slit of the monochromator by a collimating mirror. From here the beam is passed

to a sample compartment through one of the filters, which helps to eliminate unwanted

second order radiation from the diffraction grating. Upon leaving the sample compartment,

the beam is passed to the silicon photodiode detector and causes the detector to produce

an electrical signal that is displayed on the digital display.

100%T

0 Abs

Advertising