Troubleshooting – LABEC V-11D User Manual

Page 15

Advertising
background image

- 12 -

Troubleshooting

PROBLEM

Possible Cause

Solution

Power cord not connected to outlet

Plug instrument in

Dead Power outlet

Change to a different outlet

No

functioning

after power is

on .

Internal fuse melted or defective

electronic component

Change fuse or call an authorized

service engineer

Light beam blocked:

Holder misaligned

Check sample holder

Lamp is old or defective

Replace lamp

Lamp is off alignment

Readjust the lamp

Instrument

cannot

set

100%T (0A)

Defective electronic component

Call

an

authorized

service

engineer

Insufficient warm up time

warm up at least 20 minutes

Poor grounding

Check ground connection

Excessive

vibration,

strong

air

current near light source ,or strong

external light.

Improve working conditions

Lamp old or defective

Replace with a new lamp

Unstable power supply

equip a manostat to make the

voltage stable

unstable display

Defective or dirty detector or

defective electronic component

Call

an

authorized

service

engineer

Insufficient sample volume

Fill cuvette with more samples

Wrong wavelength setting

Failed to blank (0A/100%T)

Check analytical procedure and

wavelength setting.

Stray sample preparation vapors

Prepare

sample

away

from

instrument.

Use

proper

ventilation

Bubbles or particles in solution

Check sample preparation and

analytical procedure

incorrect

readings

obtained

Instrument

out

of

electronic

calibration

Call

an

authorized

service

engineer