EN 61010-1 gives in detail the conditions under which the

equipment can be operated safely (see also IP degree of

protection, “Technical data”, pg. 21).

Operating conditions

➨

This device is not approved for operation in potentially

explosive atmospheres.

Do not operate the device in

potentially explosive atmospheres.

➨

Controllers without the

”

`

”

mark on the rating plate

are

not approved for operation with dangerous or

explosive gases or with potentially explosive or

inflammable substances. Do not operate the con-

troller with dangerous or explosive gases or with

potentially explosive or inflammable substances.

➨

Controllers bearing the

”

`

”

mark on their rating

plates are approved for operation with potentially

explosive atmospheres according to their classifica-

tion

II 3G IIC T3 X according to ATEX, but they are

not approved for operation in potentially explosive

atmospheres (see section “

`

Important information:

Equipment marking (ATEX)”, pg. 19)

.

• Ensure that the materials of the controller’s wetted

parts are compatible with the substances in the vacu-

um system, see section “Technical data”, pg. 21.

Safety during operation

➨

Adopt suitable measures to prevent the release of

dangerous, toxic, explosive, corrosive, noxious or pol-

luting fluids, vapors and gases.

+

Never operate this controller if it has a damaged cord

or plug. If the controller is not working properly, has

been dropped or has fallen into water, contact your

service provider.