Read commands ”cvc 2000 – VACUUBRAND CVC 3000 detect User Manual

Page 56

Advertising
background image

page 56 of 74

Command

Operation

Response

Description

IN_PV_1

current pressure

XXXX mbar/

Torr/hPa

unit according to preselections

IN_PV_2

current frequency

XX.X Hz

pump speed

IN_CFG

device set

preselections

0XXXX

1XXXX

2XXXX

3XXXX

X0XXX

X1XXX

XX0XX

XX1XX

VACUU•LAN

continuous pumping

vacuum control without detect

vacuum control with detect

no coolant valve

coolant valve

no venting valve

venting valve

IN_CFG

device set

preselections

XXX0X

XXX1X

XXXX0

XXXX1

no automatic switch off

automatic switch off

remote operation off

remote operation on

Read commands CVC 2000

Pin assignment RS-232 C

2: RxD

5: Mass

9: +5V (Bluetooth)

3: TxD

7: RTS

4: DTR

8: CTS

A maximum of ten commands per second is possible.

Read commands and commands ”REMOTE”, ”CVC”, and ”STORE”

can always be sent. The sending of other write commands is only pos-

sible, if ”Remote on” is selected.

The commands have to be written in capital letters.

Command and parameter have to be separated by a blank.

The string is terminated with <CR> or <LF> or <CR><LF>.

The response of the controller is always terminated with <CR><LF>.

Numerical values and parameters can be written without leading ze-

ros.

The response of the controller always includes leading zeros.

Advertising