Calibration in the factory, Cleaning the pressure transducer, Control of measuring equipment – VACUUBRAND CVC 3000 detect User Manual

Page 53: Notice

Advertising
background image

page 53 of 74

Calibration in the factory

To order

DAkkS calibration of the CVC 3000 detect pressure transducer,

order number: ..............................................................................

900217

Control of measuring equipment

The

VACUUBRAND DAkkS calibration laboratory is accredited by the

Deutsche Akkreditierungsstelle GmbH (national accreditation body of the

Federal Republic of Germany) for the measurable variable

pressure in

the pressure range from 7.5*10

-4

Torr to 975 Torr (10

-3

mbar to 1300

mbar) in accordance with the general criteria for the operation of testing

laboratories defined in the DIN EN ISO/IEC 17025:2000 series of stan-

dards (accreditation number D-K-15154-01). The DAkkS is signatory to

the multilateral agreements of the European cooperation for Accredita-

tion (EA) and of the International Laboratory Accreditation Cooperation

(ILAC) for the mutual recognition of calibration certificates.

Rely on calibration in the VACUUBRAND calibration laboratory:

- To meet the requirements of the DIN ISO 9000ff and 10012 series of

standards regarding the calibration of inspection, measuring and test

equipment at specified intervals.

- To document that the vacuum gauges calibrated are traceable to na-

tional standards of the PTB (Physikalisch-Technische Bundesanstalt;

German national institute for science and technology and the highest

technical authority of the Federal Republic of Germany for the field of

metrology and certain sectors of safety engineering).

Attention: Never use a pointed or sharp-edged tool to

clean the pressure transducer.

Never touch the ceramic diaphragm of the pressure

transducer with hard objects.

Fill the measurement chamber via the vacuum connection with a sol-

vent (e.g., benzene) and allow sufficient cleaning time. Observe all

regulations concerning usage and disposal of solvents!

Cleaning the pressure transducer