Notes on configuring the controller

Time delay

• At the end of the process the coolant continues to flow according to the preset time delay. If an in-line

valve is connected, a connected pump also runs for the preset time delay with in-line valve closed for

self cleaning.

Notes on selecting the factory-set configuration

The controller VNC 2 can be adapted optimally to the specific application by choos-

ing the appropriate factory-set mode depending on the components, the application,

and the process.

Automatic valve detection:

When switching on, the controller checks the actual valve configuration. If a valve is

connected, it is monitored and operated automatically until the controller is switched

off. In-line valves and coolant valves are monitored equally.

Automatic NT VARIO detection (only with VACUUBRAND NT VARIO pumps

and VNC 2 VARIO E):

If after switching on the pump (without in-line valve) a NT VARIO-pump is detected,

the controller changes automatically to configuration ”NT VARIO” and stores this

configuration. After a restart the configuration ”NT VARIO“ is active directly.

The preselected values from last operation (e. g. for pressure, speed or switch-off

time) are stored. In case of similar operation conditions it is possible to start immedi-

ately, if the preselections are chosen appropriately.

If selecting another factory-set mode (keep key ”Esc” pressed while switching on),

the configuration for this mode is adopted, the actual configuration gets replaced.

Four factory-set modes are stored in the controller (see section „General view of

factory-set modes“, pg. 19). The individual factory-set modes contain the following

function-specific settings in the program menu:

Vacuum control (pControl): Two point control with 5 minutes time delay.
p set

: 25 mbar, ∆p: Auto, t Off: off, limit: 1060 mbar

VACUU•LAN (VACUULAN):
p set

: 25 mbar, p On: 150 mbar, t Off: 15 min

VACUU•LAN for diaphragm pumps with 8 cylinders with start-up control (3 minutes

start-up delay) (VCL 8cyl):
p set

: 25 mbar, p On: 150 mbar, t Off: 15 min

RC 5 / RC 6 management:

Attention: For factory-set mode ”RC 5 / RC 6 management” please order separate

instructions for use!

NOTICE