VACUUBRAND VNC 2 User Manual

Page 36

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999161 / 27/03/2012

page 36 of 45

Read commands compatible to ”CVC 3000”

2: RxD

7: RTS

3:

TxD

8:

CTS

4: DTR

9: +5V (Bluetooth)

5: Mass

Connector assignment

Description

unit according to preselections

maximum speed

(100% = ”HI”)

start-up pressure for VACUULAN or

two point control

delay

(00:00 = Off)

switch off pressure (”Maximum” for ”Vacuum con-

trol”, ”Minimum” for ”Continuous pumping”)

process runtime

00:00 = off

Response

XXXX mbar or

XXXX Torr or

XXXX hPa

XXX%

XXXX mbar or

XXXX Torr or

XXXX hPa

XX:XX hh:mm

XXXX mbar or

XXXX Torr or

XXXX hPa

XX:XX hh:mm

Function

Command

IN_SP_1

IN_SP_2

IN_SP_3

IN_SP_4

IN_SP_5

IN_SP_6

set vacuum

speed

start-up

pressure

delay

switch off

pressure

A maximum of ten commands per second is possible.

Read commands and commands ”REMOTE”, ”CVC” and ”STORE” can be sent always. The sending

of other write commands is only possible, if ”Remote On” is selected.

The commands have to be written in capital letters.

Command and parameter have to be separated by a blank.

The string is terminated with <CR> or <CR><LF>.

The answer of the controller is always terminated with <CR><LF>.

Numerical values and parameters can be written without leading zeros.

The answer of the controller is always with leading zeros.