• Vacuum control of a vacuum pump (with in-line valve if applicable)

Settings in the program menu:

Preset pressure (p Set): Lower pressure value for two-point control (displayed during active con-

trol by a short tip on key t).

Hysteresis (∆p): Control range of the two-point control.

Switch-off time (t Off): The duration of the process (vacuum control) can be preset. If the pump is

switched off due to surpassing of the pressure limit (Limit), the switch-off time is without effect.

Maximal pressure (Limit): Shut down of the pump when the ”Limit” value is exceeded, e.g. at

the end of filtrations or extractions (active only

after the pressure did fall below the preset pressure

(p Set) for the first time).

Suggested values for the hysteresis ∆p are stored in the controller (setting: ”Auto”) and are adapted

automatically to the preset pressure values. It is possible to adjust the parameters manually anytime.

Program menu in basic mode pControl with a NT VARIO pump

pControl
Frequen.

Frequen.
0 %

• Vacuum control with a speed controlled pump (VACUUBRAND NT VARIO)

Settings in the program menu:

Preset pressure (p Set): Exact vacuum control according to a preset pressure value.

Frequency (Frequen.): Only in case of pumps with speed control (NT VARIO). Sets the motor

speed (and thus the pumping speed) when the preset pressure is decreased manually.

Switch-off time (t Off): The duration of the process (vacuum control) can be preset. If the pump is

switched off due to surpassing of a pressure limit (Limit), the switch-off time is without effect.

Maximal pressure (Limit): Shut down of the pump when the ”Limit” value is exceeded, e.g. at

the end of filtrations or extractions (active only

after the pressure did fall below the preset pressure

(p Set) for the first time).

p in mbar

5

10

50

80

100

200

500

700

900

1000

p in mbar

(suggested values)

2

2

5

8

9

17

40

55

71

78

pControl
p Set

pControl
t Off

p Set
25mbar

t Off
off

1013mbar
Start

pControl
Limit

Limit
1060mbar

Esc