Cleaning the pressure transducer, Cleaning the pressure transducer“, pg. 40, Notice – VACUUBRAND VNC 2 User Manual

Page 40

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999161 / 27/03/2012

page 40 of 45

Cleaning the pressure transducer

The controller itself is maintenance-free.

Contamination of the pressure transducer or deposits will influence the accuracy of measurement.

Attention: Never use hard objects to clean the pressure transducer!

➨

Fill the chamber of the pressure transducer with a solvent (e. g. benzene) and allow sufficient clean-

ing time. Observe all regulations concerning usage and disposal of solvents!

➨

Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.

➨

Rinse the chamber of the pressure transducer several times with alcohol in order to remove all sol-

vent residues.

☞

Never use a spiky or sharp-edged tool to clean the pressure transducer.

➨

Allow the pressure transducer to dry.

➨

Readjust the pressure transducer if necessary.

NOTICE

Fault

Possible cause

Remedy

❏ Controller does not

respond when operating

any key. No change after

switch off/on.

✔ Contact local distributor.

Advertising