Switching to ”continuous pumping” (Cont.pump) using the function menu (see

section „Menu structure of controller“, pg. 18)

Operation of a vacuum pump using an in-line valve or direct switching of a pump

Settings in the program menu:

Switch-off time (t Off): The duration of the process control (continuous pumping) can be preset.

Limit pressure (Limit): Shut down of the pump, if the actual pressure has fallen below a preset

pressure limit ”Limit”, e.g. for drying chamber applications.

Program menu in basic mode Cont.pum when operating a valve or a pump

Cont.pum
t Off

t Off
off

Cont.pum
Limit

Limit
off