Write commands compatible to ”CVC 3000”

** If remote operation is selected or deselected, the user has to ensure that no dangerous status of

the system due to the change of the mode of operation can occur and to provide appropriate safety

measures, especially if selecting remote operation interferes with a locally operated active process.

* Pressure setting with venting is only possible in operation mode ”Vacuum control”, if a venting valve is

connected and configurated and ”Vacuum control” is started. The venting valve opens automatically,

if the actual vacuum is 10 mbar below the preset vacuum. Automatic venting becomes inactive, if

”Vacuum control” is stopped (”Esc”), if setting a pressure value using the command OUT_SP_1 or if

the operation mode is changed. Activate the command OUT_SP_V again if necessary.

X

X

X

X

Function

operation mode

set vacuum

set vacuum

with venting*

speed

start-up pressure

delay

switch off pressure

switch off time

delay

Attention: If control

is running only

switching from 1 to

2 and 2 to 1 is pos-

sible.

Description

0: VACUULAN

1: Pump down

2: Vac control

4: RC 5/6 management

Parameter

X

XXXX

XXXX

XXX

XXXX

XX:XX

XXXX

XX:XX

Command

OUT_MODE

OUT_SP_1

OUT_SP_V

OUT_SP_2

OUT_SP_3

OUT_SP_4

OUT_SP_5

OUT_SP_6

unit according to preselection

(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)

unit according to preselection

(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)

speed in %, (100% = ”HI”)

unit according to preselection

(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)

hh:mm (hours:minutes)

unit according to preselection

(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)

hh:mm (hours:minutes)

00:00 = off

0 remote off

1 remote on

2 CVC 2000 commands

3 CVC 3000 commands

0 venting valve closed

1 venting valve open

store settings permanently

1 stop

2 stop with adoption of the set vacuum

START

STOP

REMOTE**

CVC

OUT_VENT

STORE