Readjustment

The vacuum gauge was adjusted using factory standards, which are traceable through regular calibra-

tion in an accredited laboratory (DAkkS calibration laboratory) to the national standard. Depending on

the process and/or accuracy requirements, check the adjustment from time to time and readjust if nec-

essary. For readjustment, the device has to be adjusted both at atmospheric pressure as well as under

vacuum.

Adjustment at atmospheric pressure

Ventilate the controller and/or the vacuum system. Make sure

that the vacuum connection at the controller is at atmospheric

pressure.

Select program ”Cal.Sens” at controller.

Use keys s and t to adjust the display to the actual local

atmospheric pressure.

Confirm value with key ”Enter”.

Note: To determine the actual atmospheric pressure, use an accurate barometer or get accurate read-

ing from the weather service, the next airport.......(take into account the difference in altitude between e.

g. airport and laboratory).

Adjustment under vacuum

Evacuate the controller via the vacuum connection to a pres-

sure < 0.5 mbar (e. g. by applying a good rotary vane pump).

Select program ”Cal.Sens” at controller.

The display is set to zero automatically.

Confirm value with key ”Enter”.

Note: Adjustment under vacuum with an actual pressure higher than 0.5 mbar reduces the accuracy of

the measurement. If the pressure is significantly higher than 0.5 mbar, adjustment to a reference pres-

sure is recommended.

Adjustment at a reference pressure

Instead of adjustment under vacuum to a pressure < 0.5 mbar,

adjustment to a reference pressure within the range of 0 .....

20 mbar is possible.

Evacuate the controller via the vacuum connection to a refer-

ence pressure within the range 0 .... 20 mbar.

Select program ”Cal.Sens” at controller.

The display is set to zero automatically.

Use keys s and t to adjust the display to the actual refer-

ence pressure at the vacuum connection within the range

of 0 .... 20 mbar.

Confirm value with key ”Enter”.

Note: The accuracy of the value of the reference pressure will directly affect the accuracy of the adjust-

ment. If the nominal ultimate vacuum of a diaphragm pump is used as reference vacuum, the accuracy

of the controller might be doubtful. The diaphragm pump may not achieve the specified value (due to

condensate, poor state, failure of valves or the diaphragm).

1013mbar

0mbar

0....20mbar