VACUUBRAND VNC 2 E User Manual

Page 32

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999199 / 28/11/2012

page 32 of 41

Read commands compatible to ”CVC 3000”

2: RxD

7: RTS

3:

TxD

8:

CTS

4: DTR

9: +5V (Bluetooth)

5: Mass

Connector assignment

Description

unit according to preselections

maximum speed

(100% = ”HI”)

start-up pressure for VACUULAN or

two point control

delay

(00:00 = Off)

switch off pressure (”Maximum” for ”Vacuum con-

trol”, ”Minimum” for ”Continuous pumping”)

process runtime

00:00 = off

Response

XXXX mbar or

XXXX Torr or

XXXX hPa

XXX%

XXXX mbar or

XXXX Torr or

XXXX hPa

XX:XX hh:mm

XXXX mbar or

XXXX Torr or

XXXX hPa

XX:XX hh:mm

Function

Command

IN_SP_1

IN_SP_2

IN_SP_3

IN_SP_4

IN_SP_5

IN_SP_6

set vacuum

speed

start-up

pressure

delay

switch off

pressure

➨

A maximum of ten commands per second is possible.

➨

Read commands and commands ”REMOTE”, ”CVC” and ”STORE” can be sent always. The sending

of other write commands is only possible, if ”Remote On” is selected.

➨

The commands have to be written in capital letters.

➨

Command and parameter have to be separated by a blank.

➨

The string is terminated with <CR> or <CR><LF>.

➨

The answer of the controller is always terminated with <CR><LF>.

➨

Numerical values and parameters can be written without leading zeros.

➨

The answer of the controller is always with leading zeros.