3 converted data representation – ADLINK ACL-6128A User Manual
Page 29

Low-Level Programming
• 21
3.3 Converted Data Representation
Two analog output range alternatives are provided by the ACL-6128A: Unipolar
and Bipolar. The numbering of the converted data with have different
presentations for different output ranges.
3.3.1 Unipolar
Numbering
0
4095
2048
0000 0000 0000
1000 0000 0000
1111 1111 1111
V.OUT=-Vref*(0/4095)
V.OUT=-Vref*(2048/4095)
V.OUT=-Vref*(4095/4095)
Example:
Converted
Data
=2047
Binary Code
=0111 1111 1111
Vref
=-5V
V.OUT
=-(-5 V) * ( 2047/4095) = 2.499 V
3.3.2 Bipolar
Numbering
-2048
2047
0
1000 0000 0000
1111 1111 1111
V.OUT=-Vref*(-2048/2047)
V.OUT=-Vref*(0/2047)
= 0V
V.OUT=-Vref*(2047/2047)
0000 0000 0000
Example:
Converted
Data
=500
Binary Code
=1001 1111 0100
Vref
=-5V
V.OUT
=-(-5 V) * ( 500/2047) = 1.221 V