EXFO FIP-425B/435B Fiber Inspection Probe and ConnectorMax2 Mobile User Manual
Page 80

Troubleshooting
72
FIP-400B
Solving Common Problems
An APC fiber is
connected to an
FIP-435B probe, the blue
LED is blinking and the
motor is not running
When the fiber connector is detected, this will then initiate
the auto-focus sequence.
Refresh rate is very low
Ensure that the power saving mode is deactivated.
On your smart device, close the applications that are not
in use.
Reduce the number of probes operating in the vicinity.
Bring the probe and the smart device closer together.
The FIP status LED blinks
red for 2 seconds in Live
Video mode and turns
back to blue (auto focus
timeout)
Capture the image again.
The FIP LED blinks red
for 2 seconds in Capture
mode and no analysis
results are available
There was an analysis error. Repeat the inspection process.
The user interface
displayed in the
application is truncated.
Change font size in the smart device settings to facilitate
reading comprehension.
Problem
Solution