Verifying the Intensity of the Source

Verifying the Intensity of the Source

The intensity meters allow you to make sure a source is connected to the
other end of the link and that this source corresponds to the band you are
testing.

To verify the intensity of the source:
On the Parameters tab, consult the RF Intensity Meters indicators. Green
check marks indicate an appropriate intensity. Up or down arrows indicate
that the intensity is respectively too low or too high. You can also check DC
intensity from the DC Intensity Meter.

RF Intensity

Meter indicates...

Possible causes/solution

Power too low

(Signal is too

weak to be

measured)

³

Source is not active.

³

Source is active but in the wrong bandwidth.

³

Connected source is not an FLS-5800 CD/PMD Analyzer Source.

³

Source is defective.

³

Discontinuities in the optical link.

³

High-frequency modulation of the optical signal has been overly

reduced because of a saturated EDFA.

Power too high

(Signal is too

strong to be

measured)

³

Use the High Power port instead of the Low Power port.

³

Use an attenuator with a null dispersion to take your

measurement. For more information, please contact EXFO.