EXFO FTB-5800 Chromatic Dispersion Analyzer for FTB-400 User Manual

Page 84

Advertising
background image

Troubleshooting

78

FTB-5800

Solving Common Problems

Warning

Consequences

Recommended Solution

Source power is low.

Some points will be

grayed out on the graph.

³

Check if you have the right

input port.

³

Check if the fiber link is

damaged.

³

Check for irregularities in the

fiber.

³

Use the intensity meter to

check the source’s signal.

³

Make sure your fiber link is not

too long.

The power of the source

is too high.

Some points will be

grayed out on the graph.

³

Use the intensity meter to

check the source signal.

³

Check for irregularities in the

fiber.

Module temperature is

very high.

The accuracy of your

measurements can be

affected.

Use the system within the

recommended operation

temperature.

a

a.

See Technical Specifications on page 87.

Module temperature is

very low.

The accuracy of your

measurements can be
affected.

Use the system within the

recommended operation
temperature.

a