Operating Your Chromatic Dispersion Analyzer

28

FTB-5800

Performing Reference Measurements

Performing Reference Measurements

I

MPORTANT

To ensure the best possible results, EXFO recommends performing
reference measurements for wavelength and for chromatic
dispersion. Furthermore, you must perform a wavelength reference
BEFORE performing the chromatic dispersion reference.

Skipping wavelength reference may lead to inaccurate results. You
cannot perform both references in a single step.

Reference

type

Recommended reference

frequency

Remarks

Wavelength Before each test session

and every time the

application recommends it.

³

Adjusts the source wavelength precisely.

³

The application will advise you to take a

wavelength reference whenever the

need arises (after selecting a different

source type, after module initialization,

after a certain period of time or when

temperature varies), but never during

acquisitions.
When references are necessary, a

message will appear in the status bar,

recommending a new reference.

Chromatic

dispersion

Once every month and

every time the source or

patchcord is changed.

³

Used to cancel out source, analyzer and

patchcord dispersion.

It will be taken according to the sources

you have selected

(C-source; C-reference).

³

Important for low-chromatic-dispersion

measurement.