Source:power:limit:step, Source:sweep:mode – EXFO FLS-2600B Tunable Laser Source User Manual
Page 105

Remote Control
Tunable Laser Source
97
Specific Commands
SOURce:POWer:LIMit:STEP?
Description This query returns the minimum output power step (in dBm)
that can be used when changing the source output power with
the SOUR:POW command.
Syntax
SOUR:POW:LIM:STEP?
Response
The minimum output power step in ± 99.99 dBm format.
See also
SOUR:POW:LIM:HIGH?, and SOUR:POW:LIM:LOW?
SOURce:SWEep:MODE
Description This command allows you to select the sweep mode.
Syntax
SOUR:SWE:MODE<wsp><mode>
Parameters The <mode> parameter is a boolean value representing the
mode state:
0: Continuous mode
1: Step-by-Step Mode
Example
SOUR:SWE:MODE 1
Note
This command cannot be performed during a sweep.
See also
SOUR:SWE:MODE?, SOUR:SWE:CENT, SOUR:SWE:COUN,
SOUR;SWE:COUN?, SOUR:SWE:REP, SOUR:SWE:REP?,
SOUR:SWE:REV and SOUR:SWE:REV?