Adjusting your unit according to wavelength – EXFO FLS-2600B Tunable Laser Source User Manual
Page 62

Maintenance
54
FLS-2600B
Adjusting Your Unit According to Wavelength
Adjusting Your Unit According to Wavelength
It is possible to introduce an offset in your Tunable Laser Source to correct
a wavelength deviation as measured by a reference wavelength meter.
Note: A calibrated wavelength meter is required to perform a user calibration on
your unit.
To adjust your module at a certain wavelength:
1. Turn your Tunable Laser Source on.
2. Press Setup.
3. Turn the selection dial until ADJUST is displayed. The default
pre-selected wavelength appears.
4. Press ENTER. The Edit marker starts blinking in the lower part of the
display.
5. Connect your FLS-2600B to a calibrated wavelength meter.
6. Turn the selection dial until you reach the wavelength value read on
the wavelength meter, then press ENTER to validate it.
Note: The Fine-tune feature is not available in this submenu.
Edit
nm