Programming commands, Source:sweep:continuous:program – EXFO FLS-2600B Tunable Laser Source User Manual

Page 126

Advertising
background image

Remote Control

118

FLS-2600B

Programming Commands

Programming Commands

SOURce:SWEep:CONTinuous:PROGram

Description This function is used to set the parameters for a continuous

sweep program. During a sweep program, the application

sweeps continuously between the minimum (start) and

maximum (stop) wavelengths.

Syntax

SOUR:SWE:CONT:PROG<wsp>[<start>],[<stop>],

[<speed>],[<coun>],[<rev>],[<trig>]

Parameters The <start> and <stop> parameters represent the start and stop

wavelengths for the sweep. The current spectral measurement

unit applies (nm or THz). The format must be 9999.999 nm or

999.9999 THz.
The <speed> parameter is used to set the speed of the sweep.

The format must be 99.999.
The <coun> parameter is the number of passes. The format

must be an integer from 1 to 100 or 0 for continuous.
The <rev> parameter indicates whether the reverse option is on

or off. The format must be 1|0.
The <trig> parameter indicates whether the trigger option is on

or off. The format must be 1|0.