Hanna Instruments HI 98180 User Manual

Page 16

Advertising
background image

16

pppppH CALIBRATION

H CALIBRATION

H CALIBRATION

H CALIBRATION

H CALIBRATION

It is recommended to calibrate the instrument frequently, especially if high
accuracy is required.
The pH range should be recalibrated:
• Whenever the pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
• When calibration alarm time out is expired -

“CAL DUE” blinks (if feature

is enabled in SETUP).

• If

“Outside Cal Range” message blinks during pH measurement (the

measurement range is not covered by current calibration, if feature is
enabled in SETUP).

PROCEDURE
All the familly instruments offers a choice of 7 memorized buffers (1.68, 4.01,
6.86, 7.01, 9.18, 10.01 and 12.45 pH).

HI 98182/3/4/5 allow the user to set

up to 5 custom buffers. The set custom buffers are the buffer values at 25 ºC.
When a custom buffer is selected during calibration, the

“Custom” functional

key is displayed on the LCD. Press

Custom key in order to enter custom buffer

changing mode. Use

ARROW keys to change the value in a ±1.00 pH

window, in according with the temperature reading and then

Accept. Press

ESC to leave custom buffers value unchanged.
For accurate pH measurements, it is recommended to perform a calibration in
maximum allowed points. However, at least a two-point calibration is sug-
gested.
The instrument will automatically skip the buffers used during calibration and
the buffers which are in a ±0.2 pH window, around one of the calibrated buffers.
• Pour small quantities of selected buffer solutions into clean beakers. For

accurate calibration use two beakers for each buffer solution, the first one for
rinsing the electrode and the second one for calibration.

• Remove the protective cap and rinse the electrode with some of the buffer

solution to be used for the first calibration point.

FIVE-POINT CALIBRATION

FIVE-POINT CALIBRATION

FIVE-POINT CALIBRATION

FIVE-POINT CALIBRATION

FIVE-POINT CALIBRATION

• Immerse the pH electrode and the temperature

probe approximately 4 cm (1½”) into a buffer
solution of your choice (pH 1.68, 4.01, 6.86, 7.01,
9.18, 10.01, 12.45 or a custom buffer) and stir
gently. The temperature probe should be close to
the pH electrode.