• When the reading is stable and close to the selected buffer, the

CFM

functional key is displayed.

• Press

CFM to confirm calibration.

• The calibrated value and the third expected buffer value will be dis-

played.

• After the second calibration point is confirmed, immerse the pH electrode

and the temperature probe approximately 4 cm (1½”) into the third
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.

• If necessary, press the

ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected buffer, the

CFM

functional key is displayed.

• Press

CFM to confirm calibration.

• The calibrated value and the fourth expected value will be displayed.