displayed.

• Set

1800.0 mV on the simulator.

• When the reading is stable and close to the selected calibration point, the

CFM functional key is displayed.

• Press

CFM to confirm. The instrument returns to calibration screen.

• Press

ESC to return to measurement mode.

Notes:•If the reading is not close to the selected calibration point, “WRONG”

tag will blink. Verify calibration condition or contact your vendor if
you cannot calibrate.

•Press CAL or

ESC in any moment of the calibration process. The

instrument will return in the measurement mode.