Range selection, Measurement procedure – Hanna Instruments HI 96101 User Manual

Page 16

Advertising
background image

16

• Press RANGE/GLP/ to change the range.

The range can be changed any time when the
instrument is in

measurement mode

.

The selected range is memorized and the selection remaines unchanged at power off or when
the battery is removed.

The HI 96101 can measure pH when range P1 is selected, Free Chlorine when range P2 is
selected, Total Chlorine when range P3, Cyanuric Acid when range P4, Iodine when range P5,
Bromine when range P6, Iron LR when range P7 are selected. To change the active range follow
the procedure:
• Turn the meter on by pressing ON/OFF. The

display briefly shows all tags on.

• After startup, the range identification number is

displayed on the secondary LCD as P1, P2, P3,
P4, P5, P6 or P7.

RANGE SELECTION

Code

Parameter

P1

pH

P2

Free Chlorine

P3

Total Chlorine

P4

Cyanuric Acid

P5

Iodine

P6

Bromine

P7

Iron LR

MEASUREMENT PROCEDURE

To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.
Important note:

Free and Total Chlorine have to be measured separately with fresh unreacted

samples if both values are requested.

• Turn the meter on by pressing ON/OFF.

The display briefly shows all tags on.