• Place the cuvette into the cuvette holder and

ensure that the notch on the cap is positioned
securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the countdown

prior to measurement. An audible “beep”
indicates the end of the countdown period.

• Alternatively, wait for 30 seconds then just

press READ/ /TIMER. In both cases, the
lamp, cuvette and detector icons will appear
on the display, depending on the measurement
phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
iron LR on the LCD.

INTERFERENCES

Interference may be caused by:
Cadmium above 4.0 mg/L; Chromium

3+

above 0.25 mg/L; Chromium

6+

above 1.2 mg/L;

Cobalt above 0.05 mg/L; Copper above 0.6 mg/L; Cyanide above 2.8 mg/L; Manganese above
50.0 mg/L; Mercury above 0.4 mg/L; Molybdenum above 4.0 mg/L; Nickel above 1.0 mg/L;
Nitrite ion above 0.8 mg/L.
Sample pH should be between 3 and 4 to avoid developed color to fade or turbidity formation.