G l p – Hanna Instruments HI 96101 User Manual

Page 32

Advertising
background image

32

To display the calibration date:

• Press and hold for three seconds

RANGE/GLP/

to enter

GLP mode

. For

the selected range the calibration month
and day will appear on the main display
and the year on the secondary display.

• If no calibration was performed, the

factory calibration message, “F.CAL” will
appear on the main display and the
instrument returns to

measurement mode

after three seconds.

It is possible to delete the calibration and
restore factory calibration.

• Press and hold for three seconds

RANGE/GLP/

to enter

GLP mode

for

the selected range.

• Press READ/ /TIMER to enter in the

factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.

• Press ZERO/CFM to restore the factory

calibration or press RANGE/GLP/ again
to abort factory calibration restore.

• The instrument briefly indicates “donE”

upon restoration of factory calibration prior
to returning to

measurement mode

.

In the GLP mode, the last user calibration date can be verified and the factory calibration can be
restored.

LAST CALIBRATION DATE

FACTORY CALIBRATION RESTORE

G L P