Hanna Instruments HI 96101 User Manual

Page 22

Advertising
background image

22

• Fill a beaker up to the 25 mL mark with

the sample, add the content of one packet
of HI 93722-0 reagent and swirl gently
to mix. This is the reacted sample.

• Fill a second cuvette with 10 mL of the

reacted sample, up to the mark, and
replace the cap.

• Place the second cuvette into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.

• Alternatively, wait for 45 seconds then

just press READ/ /TIMER. In both
cases, the lamp, cuvette and detector
icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the

instrument directly displays concentration
in mg/L of cyanuric acid on the LCD.