PH CALIBRATION FOR HI9318

Procedure

••

•• Switch the instrument on after connecting the

electrode and temperature probe.

••

•• Immerse the pH electrode and the ATC probe

into pH 7.01 buffer solution. Shake briefly and
wait 1-2 minutes for thermal equilibrium.

Note: the electrode should be submerged approxi-

mately 4 cm (1½") into the solution. The temperature probe
should be positioned as close to the pH electrode as possible.

••

•• Press the CAL key.

The printer will then print
step by step instructions for
the calibration procedure.

At the same time, the pH symbol will flash to indicate that
calibration is in progress. The pH value at the measured (or
manually) adjusted temperature will be displayed (see page 26).
If "Err4" appears on the display it means the electrode is in the
wrong buffer. Check to make sure that the electrode is in the pH
7.01 beaker and wait for the "Err4" to disappear.

••

•• After approximately 30 seconds, press the CON key to confirm the

calibration.
The printer will print further instructions for the calibration procedure.

CAL

ON

4 cm

(1½

")

HI 7007

Dip the electro-

de in the pH 7

solution; wait

30 sec and then

press CON

ºC

pH

ºC

pH

°C

°C

Now dip the e-

lectrode in 2nd

solution, wait

30 sec and then

press CON

CON