TROUBLESHOOTING GUIDE

TROUBLESHOOTING GUIDE

TROUBLESHOOTING GUIDE

TROUBLESHOOTING GUIDE

TROUBLESHOOTING GUIDE

Symptoms are:

The instrument does

not work with the ATC

probe.

The meter fails to

calibrate or gives er-

roneous results.

The meter is slow in

responding or gives

erroneous readouts

The meter does not

accept the 2nd buffer

solution

The reading drifts

"Err1" on the display

"Err2" on the display

(for HI9017 and

HI9318 only)

"Err3" on the display

"Err4" on the display

"Err5" on the display

The problem is:

A defective ATC probe

A defective pH elec-

trode

The electrode is not

working or the dia-

phragm is clogged

A defective pH elec-

trode

A defective pH electrode

Out of range on the

pH scale

Out of range in mV

Out of range in tem-

perature

Erroneous buffer so-

lution used for offset

calibration

Defective electrode

Erroneous buffer so-

lution used for slope

calibration

Defective electrode

Possible Solutions:

Replace the probe.

Replace the electrode

and repeat the pH cali-

bration

Leave the electrode in a

storage solution after

cleaning the junction. If

the problem is not solved,

replace the electrode and re-

calibrate the pH scale.

Try the cleaning procedure.

If this does not work, re-

place the electrode.

Replace the electrode.

a) Verify the pH electrode

is connected.

b) Make sure the

sample's pH is included

between 0 and 14

c) Check the level of the

electrode's filling electro-

lyte and the state of the

electrode itself.

a) Verify the pH electrode

is connected.

b) Make sure the mV is

within ±1999 range.

Make sure the temperature

is within -10 and 125°C

Make sure that the buffer

solution used is pH 7 and

replace if necessary.

Replace the electrode

and re-calibrate pH

By sure the correct pH

solution is being used for

slope calibration.

Check the electrode and

replace if necessary.

Note: when the temperature display is showing "Err3" the pH

display is always showing "Err1".