RELATIVE mV MEASUREMENTS (HI 111 and HI 113)
• Press RANGE until “rEL” message will be displayed on the secondary LCD line

for about one second and “mV” tag will blink. After one second the

temperature will be displayed on the secondary LCD.

The reading displayed by the instrument is equal to the difference between

the current mV input value and relative mV offset established in the relative

mV calibration.
TEMPERATURE MEASUREMENTS
Connect the HI 7669/2Wtemperature probe to the appropriate

socket and turn the instrument on.

Immerse the temperatureprobeinto the sample and allow

the reading on the secondary LCD to stabilize.
MEMORY FUNCTION
Press and hold down MEM. The last reading will appear

on the LCD along with “MEM” tag until MEM is released.

Press MR and the previously memorized reading will be

displayed on the LCD with “MEM” tag.

pH CALIBRATION

Calibrate the instrument frequently, especially if high accuracy is required.

The instrument should be recalibrated:

• Whenever the pH electrode is replaced.

• At least once a week.

• After testing aggressive chemicals.

• If “CAL” “INTV” tags are blinking during measurement.

Every time you calibrate the instrument use fresh buffers and perform an

electrode Cleaning Procedure (see page 30).