Hanna Instruments HI 110 User Manual

Page 12

Advertising
background image

12

THREE- (HI 110, HI 111) / FIVE-POINT (HI 112,

HI 113) CALIBRATION
• Immerse the pH electrode and the temperature probe

approximately 4 cm (1½”) into a buffer solution and

stir gently. The temperature probe should be close to

the pH electrode.

• Press CAL. The “CAL” and “

BUF

” tags will appear

and the “7.01” buffer will be displayed on the

secondary LCD.

• If necessary, press the ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is

stable.

• When the reading is stable and close to the selected buffer, the

“READY” tag will be displayed and the “CFM” tag will blink.

• Press CFM to confirm calibration.

• The calibrated value will be displayed on the primary LCD

and the second expected buffer value on the secondary

LCD.

• After the first calibration point is confirmed, immerse the pH electrode and the

temperature probe approximately 4 cm (1½”) into the second buffer solution

and stir gently. The temperature probe should be close to the pH electrode.

• If necessary, press the ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is

stable.

• When the reading is stable and close to the selected buffer, the

“READY” tag will be displayed and the “CFM” tag will blink.

• Press CFM to confirm calibration.

• The calibrated value is then displayed on the primary

LCD and and the third expected buffer value on the

secondary LCD.

• After the second calibration point is confirmed, immerse the pH electrode

and the temperature probe approximately 4 cm (1½”) into the next

buffer solution and stir gently. The temperature probe should be close to

the pH electrode.

BUFFER pH

2

BUF

pH

CAL

BUFFER pH

1

BUF

pH

CAL