mV CALIBRATION

(for technical personnel only)

All the instruments are factory calibrated for mV.

Hanna’s ORP electrodes are interchangeable and no mV calibration is

needed when they are replaced.

If the mV measurements are inaccurate, mV recalibration should be performed.

For an accurate recalibration, contact your dealer or the nearest Hanna

Customer Service Center, or follow the instructions below.

A two or three-point calibration can be performed at 0.0 mV, 600.0 mV and

1800.0 mV.

• Attach to the BNC connector a mV simulator with an accuracy of ±0.1 mV.

• With the instrument off, press and hold down the MR&CFM keys, then

power on the instrument. The “CAL” tag will appear and the secondary

LCD will show 0.0 mV.

• Set 0.0 mV on the simulator.

When the reading is stable and close to the selected calibration point,

“READY” tag will appear and “CFM” tag will blink.

• Press CFM to confirm. The secondary LCD will display 600 mV.

• Set 600.0 mV on the simulator.

When the reading is stable and close to the selected calibration point,

“READY” tag will appear and “CFM” tag will blink.

• Press CFM to confirm. The secondary LCD will display 1800 mV.

• Set 1800.0 mV on the simulator.

When the reading is stable and close to the selected calibration point,

“READY” tag will appear and “CFM” tag will blink.

• Press CFM to confirm. The instrument returns to measurement mode.
Note: • If the reading is not close to the selected calibration point,

“WRONG” tag will blink. Verify calibration condition or contact

your vendor if you can not calibrate.

• Press CAL in any moment of the calibration process. The instrument

will return to measurement mode. If calibration process is stopped

after 600 mV is confirmed, the 600 mV range is calibrated and

calibration parameters are memorized.