Hanna Instruments HI 110 User Manual

Page 11

Advertising
background image

11

PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible, use

plastic or glass beakers to minimize any EMC interferences.

For accurate calibration and to minimize cross-contamination, use two beakers

for each buffer solution. One for rinsing the electrode and one for calibration.

If you are measuring in the acidic range, use pH 7.01 or 6.86 as first buffer and

pH 4.01 or 1.68 (HI 112, HI 113) as second buffer. If you are measuring in the

alkaline range, use pH 7.01 or 6.86 as first buffer and pH 10.01, 9.18 or 12.45

(HI 112, HI 113) as second buffer.
PROCEDURE
Calibration can be performed in three (HI 110, HI 111) or five (HI 112, HI 113)

calibration points.

For accurate measurements, it is recommended to perform a three (HI 110, HI

111) or five-point calibration (HI 112, HI 113). However, at least a two-point

calibration is suggested.

The calibration buffer can be selected from the calibration buffer list that

includes the defined custom buffers in SETUP menu and the memorized

standard buffers:

• pH 4.01, 6.86, 7.01, 9.18, 10.01 (HI 110, HI 111)

• pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 (HI 112, HI 113)

The custom buffers are a special option that allows the user to calibrate in a

buffer solution different from a standard one. Up to two custom buffers can be

set in SETUP menu (see page 17). Each custom buffer value can be changed

in a ±1.0 pH window around the set value, during calibration, when it is

selected; the “BUFFER pH” tag will blink.
The instruments will automatically skip the buffer used during calibration and

the buffers which are in a ±0.2 pH window, around one of the calibrated

buffers.
All new calibrations will override existing stored calibration data in a ±0.2

pH window. The slopes adjacent to the new points will be reevaluated.

If the new calibration point has no correspondence in the existing stored

calibration data, it is added to it if this is not full (3 or 5 points), or the

instrument will ask which buffer will be replaced by the current buffer.

If at least a two-point calibration has been performed and an offset correction of

the electrode is wanted, keeping unchanged the existing slopes, perform a one-

point calibration with “OFFS” option selected in SETUP menu. If “Pnt” option is

selected, the slopes adjacent to the calibration points will be reevaluated.

Advertising
This manual is related to the following products: