Hanna Instruments HI 110 User Manual

Page 18

Advertising
background image

18

CALIBRATION ALARM TIME-OUT SET
Press CAL when the calibration time-out is displayed. Calibration time-out

(OFF or 1 to 7 days) will start blinking.

Press the ARROWkeys to change the calibration time-out value.

Press CFM to save the modified calibration time-out value.

Press CAL to escape without saving.

FIRST CUSTOM BUFFER SET
Press CAL when custom buffer 1 is displayed. The custom buffer (disabled – “no

cb1” or 0 to 16 pH) will start blinking.

Press the ARROW keys to change the custom buffer value.

Press CFM to save the modified custom buffer value.

Press CAL to escape without saving.

SECOND CUSTOM BUFFER SET
Press CAL when custom buffer 2 is displayed. The custom buffer (disabled – “no

cb2” or 0 to 16 pH) will start blinking.

Press the ARROW keys to change the custom buffer value.

Press CFM to save the modified custom buffer value.

Press CAL to escape without saving.

ONE-POINT CALIBRATION BEHAVIOUR
Press CAL when “1 Pnt” message is displayed on the secondary LCD. One of

the two options (“Pnt” or “OFFS”) will start blinking (see pH CALIBRATION

PROCEDURE for details, page 11).

Press the ARROW keys to toggle between “Pnt” and “OFFS” options.

Press CFM to save the behaviour for one-point calibration.

Press CAL to escape without saving.

CAL

BUFFER pH

CAL

BUFFER pH

INTV

CAL

CAL