Chapter 2: Prerequisites
Product Service 1-800-522-7658
7
Proper blower selection can be determined from these exhaust requirements
and the total system static pressure loss. The enclosure must be connected to
a dedicated blower, a house exhaust system or a dedicated filtered exhauster,
such as a Labconco FilterMate. When using multiple FilterMates, the airflow
must be balanced.
Labconco offers accessory remote blowers listed next and in Chapter 7.
Remote Blowers (Roof-Mounted)
A). 1/4 hp direct drive. Corrosion-resistant epoxy-coated steel housing and
wheel. Blower inlet is 6" ID. Outlet dimensions are a rectangular 4.25" x
7.38" OD. See dimensional data in Chapter 7.
CFM @ Static Pressure – Inches of H
2
O
S.P. 0.0" 0.125" 0.25" 0.50" 0.75" 0.87"
CFM
595 560 515 420 300 167
Catalog #
Description
Shipping
Wt. lbs./kg
4863500
Remote Blower, 115 volts, 60 Hz, 4.4 amps
35/16
4863501
Remote Blower, 115/230 volts, 50 Hz, 5.6/2.8 amps
35/16
7053501
Explosion-Proof Remote Blower, 115 volts, 60 Hz,
4.4 amps
40/18
B). 1/4 hp or 1/3 hp belt drive with adjustable sheave to change RPM.
Corrosion-resistant epoxy-coated steel housing and wheel. Blower inlet is
10.87" ID. Outlet dimensions are a rectangular 5.5" x 10" OD. See
dimensional data in Chapter 7.
Catalog #
HP
Description
CFM@RPM Ranges at Static Pressure (in H
2
0)
Std. EP
.25" .38" .50" .62" .75" .88" 1.00"
7068200
1/4
Remote Blower,
115V, 60 Hz,
4.4 A
540@800
410@800 350@870 390@970 430@1060
7068300
1/4
Explosion Proof
Blower 115V,
60 Hz, 4.5 A
720@950 760@1050 710@1060 620@1060
7068400
1/3
Remote Blower,
115V, 60 Hz,
6.1 A
760@1050 710@1060 620@1060 430@1060 460@1150 500@1220
7068500
1/3
Explosion Proof
Blower, 115V,
60 Hz, 6.4 A
850@1130 825@1160 790@1190 770@1220 740@1260 640@1260
If the enclosure is connected directly to a house exhaust system, an adjustable
damper (or valve) must be installed to control the airflow properly. This is
equally important when a house exhaust system is controlling multiple
balance enclosures or ventilation stations. See Chapter 7 for ordering an
accessory adjustable damper.