Chapter 6: Maintaining your XPert or XVS Enclosure

Product Service 1-800-522-7658

32

Guardian 500 Calibration Procedure

Each alarm module and enclosure/fume hood is unique and needs to be individually
calibrated in the field. The procedure for the adjustment is as follows:

1. Double check the installation to make sure that monitor and power supply are

properly installed.

2. Allow 10 minutes for the monitor to warm up once power has been connected.
3. Determine the low flow set point for your monitor. This is the value where

the monitor will first indicate a low flow condition. The red light will be on
for this value. Refer to your industrial hygiene officer for the proper low flow
set point or consult the table below.

4. Adjust your enclosure/fume hood airflow to the low flow set point as

previously determined. The exhaust flow can be lowered by adjusting the
speed control on the FilterMate or by using an adjustable damper on the
exhaust blower. Typical alarm conditions are set at face velocities of 10 to 20
feet per minute below the normal operating conditions due to supply air and
exhaust air fluctuations, as well as room air cross drafts. See note 8 if the low
airflow volume or sash opening cannot be adjusted.

5. Using a properly calibrated thermoanemometer, determine the velocity

through the face of the enclosure by taking a detailed velocity traverse.
Divide the face area into equal increments. One reading per square foot of
face area is normally recommended for an accurate traverse. Compute the
average velocity for this area.

6. If the red light alarm is on, slowly turn the adjustment screw counterclockwise

until the green light is activated. If the green light is on, slowly turn the
adjustment screw clockwise until the red light comes on. Slowly turn the
adjustment screw back until the red light is activated. It is important that
these adjustments be done in small increments, at intervals about 30 seconds
apart to allow for delayed reaction of the alarm itself. The alarm low flow set
point should now be set and the red light activated.

7. Readjust the enclosure airflow to its normal operating levels. The green light

should now be activated. Calibration is now complete.

Enclosure Operating In Flow Speed

Alarm Condition Set Point Speed

100 fpm

80 - 90 fpm

80 fpm

60 - 70 fpm

60 fpm

40 - 50 fpm

8. Note: If the low airflow volume or sash opening cannot be adjusted, then a 1/4 to

1/3 of a turn counterclockwise can be adjusted to set the airflow alarm condition
at 10-25% below normal operating levels.

Guardian 500 Alarm Activation

The audio and visual alarm will activate approximately 10-30 seconds after an alarm condition
is detected. To temporarily mute the audible alarm, press and release the test/reset button.

NOTE: After an alarm condition has been detected, the red light will stay on. The
audible alarm will remain muted until airflow returns to normal levels.

Advertising
This manual is related to the following products:

XVS Ventilation Stations 4865022, XVS Ventilation Stations 4865021, XVS Ventilation Stations 4865012, XVS Ventilation Stations 4865011, XVS Ventilation Stations 4865010, XVS Ventilation Stations 4864042, XVS Ventilation Stations 4864041, XVS Ventilation Stations 4864032, XVS Ventilation Stations 4864031, XVS Ventilation Stations 4864030, XVS Ventilation Stations 4864022, XVS Ventilation Stations 4864021, XVS Ventilation Stations 4864012, XVS Ventilation Stations 4864011, XVS Ventilation Stations 4864010, XVS Ventilation Stations 4863042, XVS Ventilation Stations 4863041, XVS Ventilation Stations 4863032, XVS Ventilation Stations 4863031, XVS Ventilation Stations 4863030, XVS Ventilation Stations 4863022, XVS Ventilation Stations 4863021, XVS Ventilation Stations 4863012, XVS Ventilation Stations 4863011, XVS Ventilation Stations 4863010, XVS Ventilation Stations 4862042, XVS Ventilation Stations 4862041, XVS Ventilation Stations 4862032, XVS Ventilation Stations 4862031, XVS Ventilation Stations 4862030, XVS Ventilation Stations 4862022, XVS Ventilation Stations 4862021, XVS Ventilation Stations 4862012, XVS Ventilation Stations 4862011, XVS Ventilation Stations 4862010, XPert Balance Enclosures 3971872, XPert Balance Enclosures 3971871, XPert Balance Enclosures 3971862, XPert Balance Enclosures 3971861, XPert Balance Enclosures 3971860, XPert Balance Enclosures 3971842, XPert Balance Enclosures 3971841, XPert Balance Enclosures 3971832, XPert Balance Enclosures 3971831, XPert Balance Enclosures 3971830, XPert Balance Enclosures 3971822, XPert Balance Enclosures 3971821, XPert Balance Enclosures 3971802, XPert Balance Enclosures 3971801, XPert Balance Enclosures 3971800, XPert Balance Enclosures 3971672, XPert Balance Enclosures 3971671, XPert Balance Enclosures 3971662, XPert Balance Enclosures 3971661, XPert Balance Enclosures 3971660, XPert Balance Enclosures 3971642, XPert Balance Enclosures 3971641, XPert Balance Enclosures 3971632, XPert Balance Enclosures 3971631, XPert Balance Enclosures 3971630, XPert Balance Enclosures 3971622, XPert Balance Enclosures 3971621, XPert Balance Enclosures 3971602, XPert Balance Enclosures 3971601, XPert Balance Enclosures 3971600, XPert Balance Enclosures 3971572, XPert Balance Enclosures 3971571, XPert Balance Enclosures 3971562, XPert Balance Enclosures 3971561, XPert Balance Enclosures 3971560, XPert Balance Enclosures 3971542, XPert Balance Enclosures 3971541, XPert Balance Enclosures 3971532, XPert Balance Enclosures 3971531, XPert Balance Enclosures 3971530, XPert Balance Enclosures 3971522, XPert Balance Enclosures 3971521, XPert Balance Enclosures 3971502, XPert Balance Enclosures 3971501, XPert Balance Enclosures 3971500, XPert Balance Enclosures 3930442, XPert Balance Enclosures 3030441, XPert Balance Enclosures 3930432, XPert Balance Enclosures 3930431, XPert Balance Enclosures 3930430, XPert Balance Enclosures 3930422, XPert Balance Enclosures 3930421, XPert Balance Enclosures 3930402, XPert Balance Enclosures 3930401, XPert Balance Enclosures 3930400, XPert Balance Enclosures 3930342, XPert Balance Enclosures 3030341, XPert Balance Enclosures 3930332, XPert Balance Enclosures 3930331, XPert Balance Enclosures 3930330, XPert Balance Enclosures 3930322, XPert Balance Enclosures 3930321, XPert Balance Enclosures 3930302, XPert Balance Enclosures 3930301, XPert Balance Enclosures 3930300, XPert Balance Enclosures 3930242, XPert Balance Enclosures 3030241, XPert Balance Enclosures 3930232, XPert Balance Enclosures 3930231, XPert Balance Enclosures 3930230, XPert Balance Enclosures 3930222, XPert Balance Enclosures 3930221, XPert Balance Enclosures 3930202, XPert Balance Enclosures 3930201, XPert Balance Enclosures 3930200