Chapter 6: Maintaining your XPert or XVS Enclosure

Product Service 1-800-522-7658

39

5. To complete the CAL CONFIGURATION, be sure to enter “DONE”.

If needed, refer to the Configuration procedure provided on the
following page for additional details.

6. To start the calibration mode, enter “CALIBRATION” mode on the

display from the SETUP menu.

7. Follow the instructions on the display and simulate the low exhaust

volume by fully opening the sash. You may measure the average face
velocity for the low calibration set point or utilize the calculated value
provided in the chart above. The average face velocity for the low set
point is accurately measured by dividing the opening of the enclosure
into equal area grids consisting of at least 9 data collection points and
measuring the velocity at the center of each grid with a calibrated
thermo anemometer. Enter the low value on the display. Be careful
not to block the opening. The low exhaust volume calibration will
take about 5 seconds.

8. Now simulate the high exhaust volume by fully closing the sash to its

normal operating position. Measure the average face velocity for the
high calibration set point to confirm that the source of constant air
volume is providing the desired face velocity for the enclosure. The
average face velocity for the high set point is accurately measured by
dividing the opening of the enclosure into equal area grids consisting
of at least 3 data collection points and measuring the velocity at the
center of each grid with a calibrated thermo anemometer. Enter the
high value set point on the display. The high value must be at least 10
fpm greater than the low value. The high exhaust volume calibration
will take about 5 seconds.

9. Be sure to enter “DONE” after successfully completing the low and

high calibration set points.

10. Once calibration is completed, go to “RUN” and hit “ENTER”. The

value should read close to the high calibration set point.

11. With the sash fully open, the Digital 1000 monitor should go into

“RED LED” low air alarm if successfully calibrated.

Advertising
This manual is related to the following products:

XVS Ventilation Stations 4865022, XVS Ventilation Stations 4865021, XVS Ventilation Stations 4865012, XVS Ventilation Stations 4865011, XVS Ventilation Stations 4865010, XVS Ventilation Stations 4864042, XVS Ventilation Stations 4864041, XVS Ventilation Stations 4864032, XVS Ventilation Stations 4864031, XVS Ventilation Stations 4864030, XVS Ventilation Stations 4864022, XVS Ventilation Stations 4864021, XVS Ventilation Stations 4864012, XVS Ventilation Stations 4864011, XVS Ventilation Stations 4864010, XVS Ventilation Stations 4863042, XVS Ventilation Stations 4863041, XVS Ventilation Stations 4863032, XVS Ventilation Stations 4863031, XVS Ventilation Stations 4863030, XVS Ventilation Stations 4863022, XVS Ventilation Stations 4863021, XVS Ventilation Stations 4863012, XVS Ventilation Stations 4863011, XVS Ventilation Stations 4863010, XVS Ventilation Stations 4862042, XVS Ventilation Stations 4862041, XVS Ventilation Stations 4862032, XVS Ventilation Stations 4862031, XVS Ventilation Stations 4862030, XVS Ventilation Stations 4862022, XVS Ventilation Stations 4862021, XVS Ventilation Stations 4862012, XVS Ventilation Stations 4862011, XVS Ventilation Stations 4862010, XPert Balance Enclosures 3971872, XPert Balance Enclosures 3971871, XPert Balance Enclosures 3971862, XPert Balance Enclosures 3971861, XPert Balance Enclosures 3971860, XPert Balance Enclosures 3971842, XPert Balance Enclosures 3971841, XPert Balance Enclosures 3971832, XPert Balance Enclosures 3971831, XPert Balance Enclosures 3971830, XPert Balance Enclosures 3971822, XPert Balance Enclosures 3971821, XPert Balance Enclosures 3971802, XPert Balance Enclosures 3971801, XPert Balance Enclosures 3971800, XPert Balance Enclosures 3971672, XPert Balance Enclosures 3971671, XPert Balance Enclosures 3971662, XPert Balance Enclosures 3971661, XPert Balance Enclosures 3971660, XPert Balance Enclosures 3971642, XPert Balance Enclosures 3971641, XPert Balance Enclosures 3971632, XPert Balance Enclosures 3971631, XPert Balance Enclosures 3971630, XPert Balance Enclosures 3971622, XPert Balance Enclosures 3971621, XPert Balance Enclosures 3971602, XPert Balance Enclosures 3971601, XPert Balance Enclosures 3971600, XPert Balance Enclosures 3971572, XPert Balance Enclosures 3971571, XPert Balance Enclosures 3971562, XPert Balance Enclosures 3971561, XPert Balance Enclosures 3971560, XPert Balance Enclosures 3971542, XPert Balance Enclosures 3971541, XPert Balance Enclosures 3971532, XPert Balance Enclosures 3971531, XPert Balance Enclosures 3971530, XPert Balance Enclosures 3971522, XPert Balance Enclosures 3971521, XPert Balance Enclosures 3971502, XPert Balance Enclosures 3971501, XPert Balance Enclosures 3971500, XPert Balance Enclosures 3930442, XPert Balance Enclosures 3030441, XPert Balance Enclosures 3930432, XPert Balance Enclosures 3930431, XPert Balance Enclosures 3930430, XPert Balance Enclosures 3930422, XPert Balance Enclosures 3930421, XPert Balance Enclosures 3930402, XPert Balance Enclosures 3930401, XPert Balance Enclosures 3930400, XPert Balance Enclosures 3930342, XPert Balance Enclosures 3030341, XPert Balance Enclosures 3930332, XPert Balance Enclosures 3930331, XPert Balance Enclosures 3930330, XPert Balance Enclosures 3930322, XPert Balance Enclosures 3930321, XPert Balance Enclosures 3930302, XPert Balance Enclosures 3930301, XPert Balance Enclosures 3930300, XPert Balance Enclosures 3930242, XPert Balance Enclosures 3030241, XPert Balance Enclosures 3930232, XPert Balance Enclosures 3930231, XPert Balance Enclosures 3930230, XPert Balance Enclosures 3930222, XPert Balance Enclosures 3930221, XPert Balance Enclosures 3930202, XPert Balance Enclosures 3930201, XPert Balance Enclosures 3930200