Labconco Biological Safety Cabinets 34269 Series User Manual

Page 12

Advertising
background image

Chapter 2: Prerequisites

Product Service 1-800-522-7658

6

Original instructions

User-Supplied Microscope Requirements –
Purifier Cell Logic Biosafety Cabinets with
Scope-Ready™ Package Only


If installing a microscope, locate the Stand-Still™ Isolation Platform in a
separate box shipped with the cabinet. The Stand-Still Isolation Platform
supports the microscope on top of the stainless steel work surface to
eliminate up to 94% of vibrations and to promote safe airflow. If using a
stereoscope with a forward extending base, the extended Stand-Still XL
Isolation Platform should be ordered from Appendix D: Accessories. Figures
from Appendix F should be used to assist your set-up.

On models with the Scope-Ready package, the Cell Logic Biosafety Cabinet
includes a clear, disposable Pure-Vu™ Seal that can be altered to
accommodate most microscope’s eyepieces. Utilize the Punch Kit
(#4027801), supplied with your cabinet, to cut the two eyepiece holes at
approximately 2.75" (69.9 mm) spacing to match the height of your scope.
As a convenience, pre-cut Pure-Vu eyepiece seals can be ordered from
Appendix D: Accessories. The pre-cut accessory Pure-Vu Seals are based off
the nominal height from the base of the microscope to eyepiece centerline
and can save you valuable time modifying the seal shipped with the Cell
Logic. If using a stereoscope with a forward extending base, one of the Pure-
Vu XL Seals should be ordered from Appendix D: Accessories. Figures from
Appendix F should be used to assist your set-up for all Pure-Vu Seals.

User Supplied Circulator Requirements – Purifier
Cell Logic Biosafety Cabinets with Temp Zone™
Package Only


The Temp Zone™ features a temperature controlled area on the work surface
where heated or chilled samples may be transferred and kept at optimum
temperature. A user-supplied circulator that continually delivers heated or
chilled water at a minimum flow rate of 0.5 GPM (1.9 LPM) with typical
volume of 6 liters is required and may be obtained from your laboratory
supply dealer. See Figure 3-6 in Chapter 3.

Quick connection fittings are included on the left side of the cabinet and are
sized for 3/8" diameter polyethylene tubing. As a convenience, an ample
supply (20'; 6 m) of 3/8" (9.5 mm) diameter tubing, insulation, and quick
connect fittings (3/8 tube x 1/8 NPT, 3/8 tube x 1/4 NPT, 3/8 tube x 3/8 NPT)
are supplied with the Temp-Zone package for connection to a water
circulator. The water circulator can be located underneath the work surface
of the cabinet or next to the cabinet on the floor.

Advertising