Labconco Biological Safety Cabinets 34269 Series User Manual

Page 60

Advertising
background image

Chapter 7: Troubleshooting

Product Service 1-800-522-7658

54

Original instructions

PROBLEM

CAUSE

CORRECTIVE ACTION

Airflow Alert goes
off and/or there is a
slight decrease in
filter life remaining
gauge

HEPA filter loading

The gauge reading steadily decreases as
the cabinet is used.

Blockage of the
return air slots or
grille

Check all return air slots and grilles to
ensure that they are not blocked or
restricted.

Blockage of the
exhaust outlet

Ensure that the exhaust outlet is not
blocked or restricted.

Blockage

or

restriction under the
work surface

Ensure that the towel catch and plenum
beneath the work surface are
unobstructed.

Contamination of
work in the cabinet

Improper technique
or procedure for the
biosafety cabinet

See “Use of the cabinet” section in the
manual.

Restriction of the
return air slots or
grille – blockage of
the exhaust outlet

Ensure that all return air slots, grilles
and the exhaust outlet are unobstructed.

External factors are
disrupting the cabinet
airflow patterns or
acting as a source of
contamination

See “Working in the Biosafety
Cabinet” section of this manual.

Cabinet is out of
adjustment/HEPA
filter(s) are defective

Have cabinet recertified.



Advertising