Initial certification – Labconco Biological Safety Cabinets 34269 Series User Manual

Page 25

Advertising
background image

Chapter 3: Getting Started

Product Service 1-800-522-7658

19

Original instructions

Initial Certification

Prior to use, a qualified certifier should certify all biosafety cabinets. Under
normal operating conditions, the biosafety cabinet should be recertified at
least annually and when relocated or serviced. The certifier should perform
the following tests, as recommended in ANSI/NSF International Standard
Number 49 in effect when the cabinet was manufactured:

• Downflow Velocity Profile Test

• Inflow Velocity Test

• Airflow Smoke Patterns

• HEPA Filter Leak Test

• Optional Canopy Alarm Test and Operation

• Vibration Test *

• Noise Level Test *

• Lighting Intensity Test *

*These tests are user comfort related tests and may be omitted at the

user’s or certifier’s discretion.


If you have any questions regarding certification agencies or need assistance
in locating one, contact Labconco’s Product Service Department at 1-800-
522-7658 or 816-333-8811.