Chapter 7: Troubleshooting

Product Service 1-800-522-7658

53

Original instructions

PROBLEM

CAUSE

CORRECTIVE ACTION

Fluorescent light not
working

Sash is closed

Open sash – Fluorescent lights will not
work with the sash closed.

Lamp(s) are defective Replace defective lamp(s)

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective

lamp

ballasts

Replace lamp ballasts.

Keypad

disconnected

or defective

Run keypad diagnostics and check
connections.

Fluorescent light is
dim or flickering

Lamp(s) are defective Replace defective lamp(s)

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective lamp
ballast

Replace lamp ballast.

UV light not
working

Sash is open

Close sash – UV light will not work
with the sash open.

Lamp is defective

Replace defective lamp.

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective

lamp

ballast

Replace lamp ballast.

Keypad

disconnected

or defective

Run keypad diagnostics and check
connections.

UV light is dim or
flickering

Lamp is defective or
is at end of operating
lifetime.

Replace defective or worn out lamp.

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective lamp
ballast

Replace lamp ballast.