Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

44

Restoring Software Default Settings

To restore the default settings to their original configuration:

1. Access the Menu, and select “Cabinet Set Up”. Press “Mute/OK”
2. In the Cabinet Set Up submenu, select “Restore Defaults”. Press

“Mute/OK”

3. Press “Mute/OK” again to restore the software defaults. Pressing

“Menu” will escape this submenu without restoring the defaults.


Table of Default Settings

UV Lamp:

Smart-Start, Continuous, Reset UV Lamp
Hourmeter, Set UV Lamp Life to 6000 hours.

Fluorescent Lamp:

Smart-Start, Reset Lamp Hourmeter, Set
Fluorescent Lamp Life to 6000 hours.

Blower: Smart-Start

Units of Measure:

Ft / Min

Security Lock:

Non-secure

Data Output Rate:

Every second via optional RS-232 Board


Calibration

This selection is for the use of qualified certifiers only to adjust the
blower speed, or the display of the filter life remaining gauge.

Diagnostic Test Operation

When the “Diagnostic Test” menu is selected, the LCD allows keypad/
hardware testing to help troubleshoot any malfunction.

Keypad Testing

When keypad testing is selected, the LCD shows when each key is
pressed, to confirm the touchpad is working. In addition, when the
“Light” and “UV Light” keys are pressed, the associated relays are
energized, so the lamps should work. The Timer button energizes an
auxiliary relay. When “Mute/OK” is pressed, the audio alarm should
work. When the “▲” key is pressed, the Airflow Alarm relay is
energized. When the “▼” key is pressed, the Blower On relay will be
energized. Finally, the sash position switches can be checked in this
mode as the LCD displays the sash position information that the
controller is receiving.