Labconco Biological Safety Cabinets 34609 Series User Manual
Page 75
![background image](/manuals/738620/75/background.png)
Product Service 1-800-522-7658
70
A
A
p
p
p
p
e
e
n
n
d
d
i
i
x
x
E
E
:
:
Q
Q
u
u
i
i
c
c
k
k
C
C
h
h
a
a
r
r
t
t
Model
34309 34409 34509 34609
Cabinet Size (in feet)
3
4
5
6
Sash
Opening
(inches) 9 9 9 9
Starting Serial #
1
0705_
0705_
0705_
0705_
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5
Nominal Average Inflow (FPM)
105+/-5
105+/-5
105+/-5
105+/-5
Supply HEPA Data
Labconco
P/N
3838400 3838401 3838402 3838403
Exhaust HEPA Data
Labconco
P/N
3838500 3838501 3838502 3838503
Motor/Blower Data
Labconco P/N
2
3832200
3832201
3832202
3832203
Motor HP
1/2
1/2
3/4
3/4
Blower Size (inches)
10 x 8
10 x 8
10 x 8
10 x 8
Fluorescent/UV Lamp Data
Fluorescent Lamps (2 each)
F25T8 TL741 F32T8 TL741
F40T8 SP41
F40T8 SP41
Light intensity (ft-candles)
3
90-100 90-100 90-100 90-100
UV Lamp
G30T8
G30T8
G30T8
G30T8
UV energy(uW-cm
2
)
4
240
240
240
240
Prefilter Data
Labconco
P/N
3858900 3858901 3858902 3858903
Postfilter Data
Labconco P/N
NA
3837201
3837202
3837203
Canopy Data
Avg. Exhaust Vol. Range (CFM)
228-251
304-333
378-416
453-498
Canopy Alarm Setpoint Vol. (CFM)
5
228-251 304-333 378-416 453-498
Nominal
Canopy
Vol.
(CFM) 323 401 480 590
Canopy Vol. Range (CFM)
311-335
385-417
458-502
569-611
Canopy Slot Area (F
2
)
0.135 0.186 0.236 0.304
Nominal Slot Velocity (FPM)
550
400
315
330
Nominal Slot Volume (CFM)
75
75
75
100
Canopy Vacuum (In. H
2
O) .15-.25
.15-.25 .15-.25 .15-.25
Slot Velocity Alarm Setpoint (FPM)
150
100
100
100
Labconco Canopy P/N
3898000
3898001
3898002
3898003
1.
The primary serial tag is on the lower outside edge of the right corner post.
The secondary serial tag is located on the front of the electronics module on
the top right side of the cabinet. The first two digits of the serial number are
the year of production, the next two are the month. The next 5 digits are the
sequence of production, and the letter following the serial number is the
revision level of the cabinet.
2.
Each motor must be programmed by Labconco for the appropriate width
cabinet.
3.
As per ANSI/NSF Std. 49 with a background of 15 ft-candles.
4.
Measured in the geometric center of the work surface.
5.
If the optional Canopy Alarm is installed