Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

45

If An Airflow Alert Activates

The most common causes of an Airflow Alert are:

• Blockage of the inlet grilles or exhaust outlet.

• Removal of the work surface or grille during operation.

Resetting the Airflow Alert System

The Airflow Alert automatically resets to normal operation once the motor
speed has stabilized.

Working In the Biosafety Cabinet

Planning

• Thoroughly understand procedures and equipment required before

beginning work.

• Arrange for minimal disruptions, such as room traffic or entry into the

room while the cabinet is in use.

Start-up

• Turn off UV light if included.

• Slowly raise the sash until the bottom of the sash aligns with the sash

indicator decal located on the left side of the work area.

• Turn on the fluorescent light and cabinet blower if the Smart-Start

features have not been activated.

• Check the air grilles for obstructions, and note the filter gauge

reading.

• Allow the cabinet to operate unobstructed for 5 minutes.

• Wash hands and arms thoroughly with germicidal soap.

• Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber

gloves. Use protective eyewear. Wear a protective mask if
appropriate.